TYSHAK II ®
Percutaneous
Transluminal
Valvuloplasty catheter

Catheter
Characteristics
The TYSHAK II®
catheter is a coaxially constructed catheter with a distally mounted
non-compliant balloon. The catheter exhibits an extremely low profile
that allows the interventionalist to use the small introducer possible.
Due to the coaxial construction, the catheter inflates and deflates
extremely fast. The TYSHAK II®
catheter provides smaller introducers and in some cases slightly
lower burst pressures than the TYSHAK®.
Super Thin Balloon
The TYSHAK II®
PTV balloon is super thin for a low deflated profile that maintains
tip flexibility. The exceptionally low profile balloon requires the
smallest introducer possible. Nominal dimensions are maintained over the
entire length of the balloon.
Maximum
Trackability
The distal shaft through
the balloon is highly flexible for exceptional maneuverability. This
combined with the pushability of the coaxial shaft provides outstanding
tracking performance.
Radiopaque marker
Platinum marker bands
facilitate reliable positioning of the balloon.
Tyshak II®
Specifications
|
Balloon
Diameter
(MM) |
Balloon
Length
(CM) |
Introducer
Size
(FR) |
Shaft
Size
(FR) |
Usable
Length
(CM) |
Guide
Wire
(Inches) |
Rated
Burst
(ATM) |
*Nominal
Pressure
(ATM) |
Catalog
No. |
|
4.0 |
2.0 |
4 |
4 |
70 |
0.021 |
6 |
4.5 |
PDC500 |
|
4.0 |
2.0 |
4 |
4 |
100 |
0.021 |
6 |
4.5 |
SN003 |
|
4.0 |
10.0 |
4 |
4 |
70 |
0.021 |
6 |
4.5 |
PDC551 |
|
5.0 |
2.0 |
4 |
4 |
70 |
0.021 |
6 |
4.5 |
PDC501 |
|
5.0 |
2.0 |
4 |
4 |
100 |
0.021 |
6 |
4.5 |
SN004 |
|
5.0 |
3.0 |
4 |
4 |
70 |
0.021 |
6 |
4.5 |
SO020 |
|
6.0 |
2.0 |
4 |
4 |
70 |
0.021 |
4 |
3.5 |
PDC502 |
|
6.0 |
2.0 |
4 |
4 |
100 |
0.021 |
4 |
3.5 |
SN005 |
|
6.0 |
3.0 |
4 |
4 |
70 |
0.021 |
4 |
3.5 |
SO021 |
|
7.0 |
2.0 |
4 |
4 |
70 |
0.021 |
4 |
3.5 |
PDC503 |
|
7.0 |
2.0 |
4 |
4 |
100 |
0.021 |
4 |
3.5 |
SN006 |
|
7.0 |
3.0 |
4 |
4 |
70 |
0.021 |
4 |
3.5 |
PDC509 |
|
7.0 |
3.0 |
4 |
4 |
100 |
0.021 |
4 |
3.5 |
SN010 |
|
8.0 |
2.0 |
4 |
4 |
70 |
0.021 |
4 |
3.5 |
PDC504 |
|
8.0 |
2.0 |
4 |
4 |
100 |
0.021 |
4 |
3.5 |
SN007 |
|
8.0 |
3.0 |
4 |
4 |
70 |
0.021 |
4 |
3.5 |
PDC510 |
|
8.0 |
3.0 |
4 |
4 |
100 |
0.021 |
4 |
3.5 |
SN011 |
|
8.0 |
4.0 |
4 |
4 |
70 |
0.021 |
4 |
3.5 |
SO029 |
|
8.0 |
5.0 |
4 |
4 |
70 |
0.021 |
4 |
3.5 |
SO032 |
|
8.0 |
6.0 |
4 |
4 |
70 |
0.021 |
4 |
3.5 |
SO030 |
|
9.0 |
2.0 |
5 |
5 |
90 |
0.025 |
3.5 |
3 |
SO007 |
|
9.0 |
3.0 |
5 |
5 |
90 |
0.025 |
3.5 |
3 |
PDC505 |
|
9.0 |
3.0 |
5 |
5 |
100 |
0.025 |
3.5 |
3 |
SN008 |
|
10.0 |
2.0 |
5 |
5 |
90 |
0.025 |
3.5 |
3 |
SO003 |
|
10.0 |
3.0 |
5 |
5 |
90 |
0.025 |
3.5 |
3 |
PDC506 |
|
10.0 |
3.0 |
5 |
5 |
100 |
0.025 |
3.5 |
3 |
SN009 |
|
10.0 |
4.0 |
5 |
5 |
90 |
0.025 |
3.5 |
3 |
SO001 |
|
10.0 |
5.0 |
5 |
5 |
90 |
0.025 |
3.5 |
3 |
SO031 |
|
10.0 |
6.0 |
6 |
6 |
100 |
0.025 |
3.5 |
3 |
SO009 |
|
11.0 |
2.0 |
5 |
5 |
90 |
0.025 |
3.5 |
3 |
SN11290 |
|
11.0 |
3.0 |
5 |
5 |
90 |
0.025 |
3.5 |
3 |
SO017 |
|
11.0 |
4.0 |
5 |
5 |
90 |
0.025 |
3.5 |
3 |
SO002 |
|
12.0 |
2.0 |
5 |
5 |
90 |
0.025 |
3.5 |
3 |
SO008 |
|
12.0 |
2.0 |
6 |
6 |
90 |
0.035 |
3.5 |
3 |
SO014 |
|
12.0 |
3.0 |
5 |
5 |
90 |
0.025 |
3.5 |
3 |
PDC507 |
|
12.0 |
4.0 |
6 |
6 |
90 |
0.035 |
3.5 |
3 |
PDC508 |
More
The devices listed have CE mark approval. Any
customer or distributor outside of the EU will be subject to their
individual country's regulations in regards to the importation and sale of
these products.
* Pressure at which balloon will
not exceed 10% of nominal size.
This publication could
include technical or other inaccuracies or typographical errors. Changes
are periodically added to the information herein; these changes will be
incorporated in the new editions of the publication. NuMED Incorporated
may make improvements and/or changes in the product(s) and/or the
program(s) described in this publication at any time. |